Identifying the optimal monopolar electrocautery output power in pedicular internal thoracic artery harvesting: 20 or 40 watts?

Yükleniyor...
Küçük Resim

Tarih

2022

Dergi Başlığı

Dergi ISSN

Cilt Başlığı

Yayıncı

Clinics Cardive Publ Pty Ltd

Erişim Hakkı

info:eu-repo/semantics/openAccess

Özet

Background: Monopolar electrocautery is an important tool for harvesting the pedicular internal thoracic artery (ITA) in cardiac surgery. The different power outputs of cautery may affect graft integrity and long-term patency. This study aimed to identify the optimal threshold of electrocautery power for ITA harvest. Methods: This prospective study included 30 patients who underwent elective coronary artery bypass surgery at the Medipol Mega University Hospital. The ITA was harvested by monopolar electrocautery after a median sternotomy. The output of cautery was adjusted at 20 W in group A and 40 W in group B. Three to 4 cm of a distal ITA sample from each patient was examined under a light microscope by two independent pathologists. Results: The ITA harvest time was longer in group A (21.2 +/- 7.5 vs 10.3 +/- 8.1 min, p < 0.001) than in group B. ITA free flow was similar in the two groups (43.6 +/- 48.7 vs 51.7 +/- 45.0 ml/min, p = 0.762). Mild to moderate injury in the endothelial and sub-endothelial sample was more frequent in the low-cautery group (p = 0.0037). Conclusion: ITA endothelial integrity was found to be better preserved with 40W electrocautery. Moreover, 20W of monop-olar electrocautery may not be safe in pedicular ITA harvesting.

Açıklama

Anahtar Kelimeler

coronary artery bypass, electrocautery, endothelial injury, internal thoracic artery

Kaynak

Cardiovascular Journal of Africa

WoS Q Değeri

Q4

Scopus Q Değeri

Q3

Cilt

33

Sayı

5

Künye