Comparing the effect of two different intraocular lenses on optical aberrations in bilaterally operated eyes for cataract
Yükleniyor...
Tarih
2013
Dergi Başlığı
Dergi ISSN
Cilt Başlığı
Yayıncı
Professional Medical Publications
Erişim Hakkı
info:eu-repo/semantics/openAccess
Özet
Objective: To assess high order and spherical aberrations results of hydrophobic acrylic AMO Sensar AR40E and hydrophobic acrylic Alcon AcrySof SA60AT intraocular lenses after implantation in cases with bilateral cataract. Methods: Cases diagnosed as bilateral cataract were included in the study and preoperative aberration measurements were recorded by using Nidek OPD SCAN-ARK 1000. Groups were created by implanting AMO Sensar AR40E to one eye of the patients, while Alcon AcrySof SA60AT into the other in a prospective and randomized manner. Aberration measurements were recorded after one and two months of surgery. Results: Overall, 40 eyes in 20 patients (11 women and 9 men) were included in the study. All patients underwent bilateral phacoemulsification surgery due to cataract. There were 20 eyes in both groups. Mean age was 62.4 (range: 31-82) years. There was no significant difference in aberrations recorded before surgery and one and two months after surgery in both groups. (p < 0.05). Conclusion: There was no difference among spherical intraocular lenses used in this study.
Açıklama
Anahtar Kelimeler
Cataract, Intraocular lens, Optical aberration
Kaynak
Pakistan Journal of Medical Sciences
WoS Q Değeri
Q4
Scopus Q Değeri
Q1
Cilt
29
Sayı
4