An effective and safe alternative to epicardial pacemaker placement for permanent pacemaker implantation in a patient with mechanical tricuspid valve: stimulation of the left ventricle through the coronary sinus
dc.contributor.author | Demir, Ahmet Duran | |
dc.contributor.author | Sen, Nihat | |
dc.contributor.author | Erbay, Ali Riza | |
dc.contributor.author | Atak, Ramazan | |
dc.date.accessioned | 2024-09-18T20:06:31Z | |
dc.date.available | 2024-09-18T20:06:31Z | |
dc.date.issued | 2011 | |
dc.department | Hatay Mustafa Kemal Üniversitesi | en_US |
dc.description.abstract | Implantation of a transvenous endocardial pacemaker is contraindicated in patients with a mechanical tricuspid valve. An epicardial left ventricular pacemaker lead was placed by a transvenous route through the coronary sinus into the lateral cardiac vein in a 58-year-old woman with mechanical aortic, mitral, and tricuspid valves, for permanent pacing due to chronic atrial fibrillation with a slow ventricular rate accompanied by syncope. This lead was then connected to a singlechamber pacemaker. The patient had no problem in the following three months. Placement of an epicardial pacing lead through the coronary sinus provides a safe and effective pacing in patients with a mechanical tricuspid valve, thus obviating major cardiac surgery. | en_US |
dc.identifier.doi | 10.5543/tkda.2011.01322 | |
dc.identifier.endpage | 247 | en_US |
dc.identifier.issn | 1016-5169 | |
dc.identifier.issue | 3 | en_US |
dc.identifier.pmid | 21532304 | en_US |
dc.identifier.scopus | 2-s2.0-79959967449 | en_US |
dc.identifier.scopusquality | Q3 | en_US |
dc.identifier.startpage | 244 | en_US |
dc.identifier.uri | https://doi.org/10.5543/tkda.2011.01322 | |
dc.identifier.uri | https://hdl.handle.net/20.500.12483/8562 | |
dc.identifier.volume | 39 | en_US |
dc.identifier.wos | WOS:000421527100012 | en_US |
dc.identifier.wosquality | N/A | en_US |
dc.indekslendigikaynak | Web of Science | en_US |
dc.indekslendigikaynak | Scopus | en_US |
dc.indekslendigikaynak | PubMed | en_US |
dc.language.iso | tr | en_US |
dc.publisher | Turkish Soc Cardiology | en_US |
dc.relation.ispartof | Turk Kardiyoloji Dernegi Arsivi-Archives of The Turkish Society of Cardiology | en_US |
dc.relation.publicationcategory | Makale - Uluslararası Hakemli Dergi - Kurum Öğretim Elemanı | en_US |
dc.rights | info:eu-repo/semantics/closedAccess | en_US |
dc.subject | Cardiac pacing | en_US |
dc.subject | artificial/methods | en_US |
dc.subject | coronary sinus | en_US |
dc.subject | electrodes | en_US |
dc.subject | implanted | en_US |
dc.subject | tricuspid valve/surgery | en_US |
dc.title | An effective and safe alternative to epicardial pacemaker placement for permanent pacemaker implantation in a patient with mechanical tricuspid valve: stimulation of the left ventricle through the coronary sinus | en_US |
dc.type | Article | en_US |