Effects of millimetric shifts in breast cancer radiotherapy on the radiation dose distribution

[ N/A ]

Tarih

2016

Dergi Başlığı

Dergi ISSN

Cilt Başlığı

Yayıncı

Asian Pacific Organization for Cancer Prevention

Erişim Hakkı

info:eu-repo/semantics/openAccess

Özet

Background: This study aimed to facilitate decision-making in cases of breast cancer radiotherapy shifts by simulating millimetric shifts and analyzing their effects on dose distribution. Methods: The study included 30 patients with left side breast cancer who were treated with three dimensional conformal radiotherapy (3D-CRT) in the Radiation Oncology Department in Hatay Public Hospital, between January 2013 and April 2015. A treatment plan shifting at three axes with six different measures was simulated. Results: The biggest difference in values was (+3mm shift) 476cGy, with a 7.7 % change for heart and 25.6% for spinal cord. The shifts in values respectively for CTV min, mean, max were -4.8%, 2.5%, 4%. The differences for lymphatic min, mean, max were 21.3%, 20.3%, -12.2%. Conclusion: The most important thing is not the treatment plan quality, but its practicality. The treatment plan must be practical and its practice must be controlled rigidly.

Açıklama

Anahtar Kelimeler

Breast cancer, Millimetric shifts, Radiation dose distribution, Radiotherapy

Kaynak

Asian Pacific Journal of Cancer Prevention

WoS Q Değeri

Scopus Q Değeri

Q3

Cilt

17

Sayı

3

Künye