Efficacy of low dose and ultra-low dose on the visibility of peri-implant fenestration and dehiscences: a computed tomography study
[ N/A ]
Tarih
2022
Dergi Başlığı
Dergi ISSN
Cilt Başlığı
Yayıncı
Int Scientific Information Inc
Erişim Hakkı
info:eu-repo/semantics/openAccess
Özet
Purpose: This study aimed to evaluate the visibility of peri-implant fenestration and dehiscences on computed tomography (CT) images taken with 2 different doses. Material and methods: The defects were created on the apical of 6 implants randomly selected from 20 titanium implants placed in the ribs, and dehiscences were created on the cervical of 8 implants. No defects were created around 6 implants. Macroscopic analysis of the implanted ribs was accepted as the gold standard. From the samples, images were taken by using both ultra-low dose (80 kVp, 50 mA, 1.25 mm slice thickness) and low dose (100 kVp, 50 mA, 1.25 mm slice thickness) protocols in CT. The images obtained were evaluated using a 5-point scale. Results: No significant difference was found between the area under the receiver operating characteristic of ultra-low dose protocol and low dose protocol in both defects based on the Wilcoxon test (p > 0.05). Conclusions: The ultra-low dose protocol could be applied by adhering to the as low as reasonably achievable principle in the diagnosis of peri-implant defects.
Açıklama
Anahtar Kelimeler
computed tomography, radiation dose, implant, fenestration, dehiscences
Kaynak
Polish Journal of Radiology
WoS Q Değeri
N/A
Scopus Q Değeri
#BAŞV!
Cilt
87