Troponin-ı ölçümünde ELFA ve CMIA yöntemlerinin karşılaştırılması

dc.contributor.advisorYönden, Zafer
dc.contributor.authorKurtgöz, Özgür Yıldırım
dc.date.accessioned2024-05-28T11:17:29Z
dc.date.available2024-05-28T11:17:29Z
dc.date.issued2013en_US
dc.departmentFakülteler, Tıp Fakültesi, Biyokimya Ana Bilim Dalıen_US
dc.description.abstractAmaç: Akut Koroner Sendrom tanı ve takibinde önemli bir yer tutan biyokimyasal kardiyak belirteçlerden kardiyak troponin-I (cTnI), spesifite ve sensitivitesi açısından diğerlerine göre daha özel bir öneme sahiptir. Bu çalışmada amaç; iki farklı troponin I ölçüm sistemini cihaz performansları açısından karşılaştırmaktır.Yöntem: ELFA (Enzim Linked Fluoro Immunassay) yöntemiyle çalışan Vidas ve CMIA (Kemilüminesans Immünassay) yöntemiyle çalışan Architect I 2000 SR cihazları arasında sistematik hatanın belirlenmesi için doğrusallık, geri kazanım, interferans, yöntem karşılaştırma deneyleri, rastgele hatanın tespit edilmesi için tekrarlanabilirlik çalışmaları yapıldı. Architect I 2000 SR cihazında çalışılan cTnI hasta sonuçları kullanılarak ROC analiz eğrisi grafiği yapıldı.Bulgular: Vidas cihaz için 0-24.8 µg/L, Architect I 2000 SR cihazı için 0-35.3 ng/ml'deki sonuçlar lineer olarak bulundu. Her iki cihazın deney içi ve deneyler arası tekrarlanabilirlik sonuçları %10 CV'den küçüktür. Architect I 2000 SR cihazında ortalama %84.2 geri kazanım, Vidas cihazında ortalama %90 geri kazanım tespit edildi. Her iki cihazın interferansı için izin verilen hata yüzdesi <%22.5'dir. Yöntem karşılaştırma deneyinde, iki cihaz arasında korelasyon katsayısı (r) 0.96 olarak bulundu. ROC analiz eğrisinde, ilk kan alımından 4-12 saat sonra çıkan sonuçların ilk başvuru anındaki (0-6. saatler) sonuçlara göre daha duyarlı ve özgül olduğu tespit edildi.Sonuç: Sonuç olarak çalışmamızda cTnI ölçümüne ait iki farklı yöntemin analitik performansları değerlendirildiğinde, iki cihazın sonuçları birbiri ile uyumlu bulundu.en_US
dc.description.abstractBackgraund and Aim: Cardiac troponin I (cTnI), an important biochemical marker for diagnosis and management of acute coronary syndrome, has particular importance than others in terms of its specificity and sensitivity. In this study, device performance of two different troponin I measurement systems was compared.Methods: Linearity, recovery, interference and method comparison experiments were carried out to be able to determine any possible systematic error between Vidas device, which is using CMIA (Chemiluminescence Immunoassay) method, and Architect I 2000 SR device, which is using ELFA (Enzyme Linked Fluoro-Immunoassay) method. Reproducibility studies were used to determine random errors. ROC curve graph was drawn by using CTnI patient results analyzed in Architect i2000.Results: The results found as 0-24.8 mg/L in Vidas and 0-35.3 ng/ml in Architect I 2000 SR, were compatible with linearity. Intra and interexperimental reproducibility results were found lower than 10% CV in both devices. Recovery was detected on average of 84.2% in Architect I 2000 SR, while that was on average of 90% in Vidas device. Error percentage which was allowed for interference of each devices was <22.5%. In the method comparison study, correlation coefficient(r) between these two devices was found to be 0.96. The results in the first 4-12 hours after blood collection were found more specific and sensitive than the results in the first admission period (0-6. hours) on ROC curve.Conclusion: The results of two devices were compatible with each other, as the analytical performance of two different methods was evaluated by measurement of cTnI.en_US
dc.identifier.endpage110en_US
dc.identifier.startpage1en_US
dc.identifier.urihttps://tez.yok.gov.tr/UlusalTezMerkezi/TezGoster?key=1zw6GvYMe-q3Hf6HR-3USxdvgo2McPQNiyqFelgFZ1x0QUuYowrFp0C9zWLQKQFk
dc.identifier.urihttps://hdl.handle.net/20.500.12483/5948
dc.identifier.yoktezid352819en_US
dc.language.isoturen_US
dc.publisherHatay Mustafa Kemal Üniversitesien_US
dc.relation.publicationcategoryTezen_US
dc.rightsinfo:eu-repo/semantics/openAccessen_US
dc.subjectBiyokimyaen_US
dc.subjectBiochemistryen_US
dc.subjectKardiyak troponin-Ien_US
dc.subjectEnzim Linked Fluoro Immunassayen_US
dc.subjectKemilüminesans Immünassay.en_US
dc.subjectCardiac troponin Ien_US
dc.subjectEnzyme Linked Fluoro-Immunoassayen_US
dc.subjectChemiluminescence Immunoassay.en_US
dc.titleTroponin-ı ölçümünde ELFA ve CMIA yöntemlerinin karşılaştırılmasıen_US
dc.title.alternativeTroponin-i measurement ELFA and CMIA comparison of methodsen_US
dc.typeSpecialist Thesisen_US